Chinese Journal of Medical Ultrasound (Electronic Edition) >
2020 , Vol. 17 >Issue 01: 11 - 16
DOI: https://doi.org/10.3877/cma.j.issn.1672-6448.2020.01.003
Preliminary application of three-dimensional shear wave elastography in diagnosis of thyroid nodules
Received date: 2019-11-27
Online published: 2020-01-01
To investigate the value of three-dimensional shear wave elastography (3D-SWE) in the differential diagnosis of benign and malignant thyroid nodules.
From December 2018 to October 2019, a total of 75 patients with 94 thyroid nodules at Sichuan Cancer Hospital were included. Ultrasound, 2D-SWE, and 3D-SWE were performed, and the maximum Young's modulus (Emax), average Young's modulus (Emean), and elastic data dispersion (Esd) were measured. All nodules were confirmed by operation and pathology. According to the pathology results, the receiver operating characteristic (ROC) curves of 2D-SWE and 3D-SWE parameters were plotted to assess their value in distinguishing benign and malignant thyroid nodules.
There were 24 benign nodules and 70 malignant nodules confirmed by pathology in this study. On all sections of 2D-SWE/3D-SWE, the Emax, Emean, and Esd of malignant nodules were significantly larger than those of benign nodules (P<0.05 for all). On the sagittal and transverse planes of 2D-SWE/3D-SWE and coronal plane of 3D-SWE, the areas under the ROC curves (AUCs) of Emax in diagnosing thyroid nodules were 0.753, 0.797, 0.79, 0.811, and 0.806, respectively, and there were no significant differences between them (P>0.05 for all); the AUCs of Emean were 0.740, 0.745, 0.771, 0.780, and 0.802, respectively, and there were no significant differences between them (P>0.05 for all). The AUC of Esd was significantly lower on the transverse plane of 2D-SWE than on the sagittal plane of 2D-SWE/3D-SWE and coronal plane of 3D-SWE (0.675 vs 0.803, 0.805, and 0.774, Z=2.540, 1.971, and 1.963, P<0.05 for all).
3D-SWE can be used to identify benign and malignant thyroid nodules by quantitative evaluation of the hardness of thyroid nodules. The diagnostic performance of Esd on the sagittal and coronal planes of 3D-SWE is significantly better than that on the transverse plane of 2D-SWE, and Esd is a useful method for differential diagnosis of benign and malignant thyroid nodules.
Ziyue Hu , Man Lu , Lu Wang , Ting Wei , Minggang Wu , Yuting Fan , Wei Yang . Preliminary application of three-dimensional shear wave elastography in diagnosis of thyroid nodules[J]. Chinese Journal of Medical Ultrasound (Electronic Edition), 2020 , 17(01) : 11 -16 . DOI: 10.3877/cma.j.issn.1672-6448.2020.01.003
1 | BercoffJ, TanterM, Fink M. Supersonic Shear Imaging: a new technique for soft tissue elasticity mapping [J]. IEEE Trans Ultrason Ferroelectr Freq Control, 2004, 51(4): 396-409. |
2 | Cho KR, Seo BK, Lee JY, et al. A comparative study of 2D and 3D ultrasonography for evaluation of solid breast masses [J]. Eur J Radiol, 2005, 54(3): 365-370. |
3 | Petzold G, Tsaknakis B, Bremer SCB, et al. Evaluation of liver stiffness by 2D-SWE in combination with non-invasive parameters as predictors for esophageal varices in patients with advanced chronic liver disease [J]. Scand J Gastroenterol, 2019, 54(3): 342-349. |
4 | 高明.头颈肿瘤学[M].北京:科学技术文献出版社, 2014: 913. |
5 | Guth S, Theune U, Aberle J, et al. Very high prevalence of thyroid nodules detected by high frequency (13 MHz) ultrasound examination [J]. Eur J Clin Invest, 2009, 39(8): 699-706. |
6 | 王涛,王学梅,张义侠, 等. 实时剪切波弹性成像鉴别甲状腺结节良恶性的定量分析[J]. 中国医学影像学杂志, 2012, 20(9): 684-687. |
7 | Ng WL, Rahmat K, Fadzli F, et al. Shearwave Elastography Increases Diagnostic Accuracy in Characterization of Breast Lesions [J]. Medicine(Baltimore), 2016, 95(12): e3146. |
8 | Liu Z, Jing H, Han X, et al. Shear wave elastography combined with the thyroid imaging reporting and data system for malignancy risk stratification in thyroid nodules [J]. Oncotarget, 2017 27, 8(26): 43406-43416. |
9 | 纪小龙, 吉米. 甲状腺病理诊断[M]. 北京: 人民军医出版社, 2011. |
10 | Tian J, Liu Q, Wang X, et al. Application of 3D and 2D quantitative shear wave elastography (SWE) to differentiate between benign and malignant breast masses [J]. Sci Rep, 2017, 7: 41216. |
11 | Dobruch K, Piotrzkowska H, et al. Ultrasound echogenicity reveals the response of breast cancer to chemotherapy [J]. Clin Imaging, 2019, 26(55): 41-46. |
12 | Sebag F, Vaillant-Lombard J, Berbis J, et al. Shear wave elastography: a new ultrasound imaging mode for the differential diagnosis of benign and malignant thyroid nodules [J]. J Clin Endocrinol Metab, 2010, 95(12): 5281-5288. |
13 | Kim HJ, Kwak MK, Choi IH, et al. Utility of shear wave elastography to detect papillary thyroid carcinoma in thyroid nodules: efficacy of the standard deviation elasticity [J]. Korean J Intern Med, 2019, 34(4): 850-857. |
14 | 郭勇,李福英,刘君. 24例亚急性肉芽肿性甲状腺炎的超声诊断[J]. 咸宁学院学报(医学版), 2007, 21(1): 68-69. |
/
〈 |
|
〉 |
Copyright © Chinese Journal of Medical Ultrasound (Electronic Edition), All Rights Reserved. Tel: 010-51322630、2632、2628 Fax: 010-51322630 E-mail: csbjb@cma.org.cn Powered by Beijing Magtech Co. Ltd |