Chinese Journal of Medical Ultrasound (Electronic Edition) ›› 2021, Vol. 18 ›› Issue (07): 681-687. doi: 10.3877/cma.j.issn.1672-6448.2021.07.012

Special Issue:

• Obstetric and Gynecologic Ultrasound • Previous Articles     Next Articles

Application of three Doppler methods in quantitative measurement of fetal atrioventricular conduction time

Mei Pan1, Bowen Zhao1(), Yankai Mao1, Xiaohui Peng1, Yuan Yang1, Bei Wang1, Xiaolu Sun1, Jun Wu2   

  1. 1. Department of Diagnostic Ultrasound & Echocardiography, Sir Run Run Shaw Hospital, Zhejiang University College of Medicine, Technical Guidance Center for Fetal Echocardiography of Zhejiang Province & Sir Run Run Shaw Institute of Clinical Medicine of Zhejiang University, Hangzhou 310016, China
    2. Department of Ultrasound, the Affiliated Hospital of Hangzhou Normal University, Hanghzou 310000, China
  • Received:2020-03-25 Online:2021-07-01 Published:2021-07-16
  • Contact: Bowen Zhao

Abstract:

Objective

To establish the normal range of fetal atrioventricular (AV) time intervals during the second and third trimesters assessed by three different pulsed-wave Doppler techniques [superior vena cava and ascending aorta (SVC/AAO) method, tissue Doppler imaging (TDI), and pulmonary artery and pulmonary vein (PA/PV) method and analyze the correlation of AV value with gestational age, fetal heart rate, and heart area.

Methods

From June to December 2019, 227 normal fetuses who underwent echocardiography at the Sir Run Run Shaw Hospital of Zhejiang University College of Medicine were selected. The gestational age was 16 to 41 weeks. They were divided into five groups according to gestational age. Pulsed-wave Doppler methods including SVC/AAO, TDI, and PA/PV were applied to measure AV, respectively. The measurement success rate and the normal range of AV values in different gestational weeks were obtained. Single-factor analysis of variance and multiple comparisons (LSD-t method) were used to analyze the differences in AV values obtained by different methods. Pearson correlation analysis was performed to examine the correlation of AV values with gestational age, fetal heart rate, and fetal heart area. Intraclass correlation coefficient (ICC) was used to evaluate the in-observer and inter-observer repeatability of each method.

Results

All the three Doppler methods achieved a high measurement success rate: SVC/AAO, 94.58%; TDI, 100%; PA/PV, 95.83%. The normal range of AV values was: AV-SVC/AAO, (115.94±6.67) ms; AV-TDI, (117.69±6.43) ms; AV-PA/PV,(123.70±6.42 )ms; the difference between them was statistically significant (t=-2.843, -12.662, -9.819; P=0.005, <0.001, <0.001). All AV values were positively correlated with gestational age (r=0.677~0.680, P<0.001) and fetal heart area (r=0.648~0.654, P<0.001), but negatively correlated with fetal heart rate (r=-0.568~-0.581, P<0.001). In-observer and inter-observer repeatability tests showed that the three Doppler methods had good repeatability (ICC>0.80).

Conclusion

All the three Doppler methods are feasible and reproducible in measuring fetal heart AV values; normal fetal AV values in the second and third trimesters are affected by gestational age, fetal heart area, and fetal heart rate, and there are differences in AV values measured by different methods. Therefore, choosing an appropriate method to measure the fetal AV time and understanding the difference between the measured values by different methods are very important in clinical work and have very important significance.

Key words: Fetal echocardiography, Atrioventricular time, Atrioventricular block, Prenatal diagnosis